Warrior Portable Warmers VS buddy lite™

Warrior & Warrior EXTREMEWarrior lite (Extra Power battery)Warrior lite (lite battery)buddy lite™
Max Delivery Rate at 4°C input [1]200 ml/min [2]180 ml/min170 ml/min50 ml/min at 10°C per user manual. See conflicting data in note [3]
Max Delivery Rate at 20°C input [1]290 ml/min270 ml/min250 ml/min80 ml/min per user manual. See note [3]
Max Battery Capacity at 4°C input [4]3.5 liters (3.3 liters at 200 ml/min) [2]3 liters1.3 litersSee note [3]
Max Battery Capacity at 20°C input [4]5 liters (per independent study, 4.8 liters at 200ml/min) [2]5.0 liters2.5 liters4.4 liters per user manual. See note [3]
Warming with Intense Push/Pull Resuscitation Techniques [5]Performance validatedPerformance validatedPerformance validatedSee note [6]
Operating weight - single battery1,790g [7]1,203g915g667g [7]
Operating weight - to warm 2.5L at 4°C input1,790g [7]1,203g1,365g (2 batteries)See note [3]
Same Disposable for Blood and IV FluidsYesYesYesYes
AC Operable OptionYesWIPWIPYes
Aluminum FreeYesYesYesYes
Disposable Unit Price$60-95 [8]$60-95 [8]$60-95 [8]$20-30 (est.) [9]
Notes:

[1] The term “Max Delivery Rate” means “the maximum flow rate in which target temperature – typically 38°C — will be reached”. Obviously, this measure relates to battery capacity as well; the assumption therefore is that a fully charged battery is used.

[2] “At an input temperature of 20°C and a flow rate of 50 mL/min, the Warrior warmed over 3.5 L to an average temperature of 37.8°C. Flow rates of 100 and 200 mL/min increased total heated volume with small effect on the output temperature (4.3 and 4.8 L at 37.1°C and 36.1°C, respectively). This behaviour persists at an input temperature of 10°C—warming 2.6, 2.9 and 3.3 L to 36.9°C , 35.9°C and 34.4°C at flow rates of 50, 100 and 200 mL/min, respectively. (Lehavi A, Yitzhak A, Jarassy R, et al. Emergency Med Journal, doi:10.1136/ emermed-2017-207112).

[3] “At an input temperature of 20°C and flow of 50 mL/min, the Buddy Lite warmed over 3 L to above 35°C, an energy transfer of more than 44 watt hour (Wh). As the flow increased to 100 mL/min, the Buddy Lite warmed intermittently for 30 min before stabilising at a peak temperature of 33°C (this behavior is demonstrated in figure 2B), heating the first 3 L by only 5°C–8°C, with a maximal temperature exceeding 35°C, achieved after more than 30 min, and lasting for a prolonged period. This translated into low initial energy transfer for the initial 3 L, but equated to heat transfer of greater than 38 Wh when more than 5 L were infused. As flow rate increased to 200 mL/min, the Buddy Lite performance pattern worsened: it consisted of short heating attempts, without stabilising at an appropriate output temperature, providing minimal energy transfer (1.5 Wh). This behaviour persisted at an input temperature of 10°C, regardless of the flow rate. The Buddy Lite was unable to sustain stable, appropriate output temperature and at flow rates of 50, 100 and 200 mL/min it heated fluid by only 1°C–2°C, with minimal heat transfer (6.7, 2.8 and 2.0 Wh, respectively)” (Lehavi A, Yitzhak A, Jarassy R, et al. Emergency Med Journal, doi:10.1136/ emermed-2017-207112).

[4] The term “Battery Capacity” means “the volume of fluids – in liters — that can be warmed with a single, fully charged battery”. Actual output is subject to battery condition, ambient temperature (typically 20°C ambient is assumed) and flow rates.

[5] E.g. hand pump; LifeFlow-like devices, etc.

[6] Based on manufacturer’s published data, and as reflected in the independent study results mentioned above, warming efficiency may not suffice for very intense bolus flows. Further, the cassette’s resiliency to sustain the extreme pressure that is often generated by intense push-pull resuscitation techniques need to be validated. It is recommended to consult with the manufacturer.

[7] Based on an independent study (Lehavi A, Yitzhak A, Jarassy R, et al. Emergency Med Journal, doi:10.1136, emermed-2017-207112)

[8] Rounded. Coming soon – lower-cost option ($15-30)

[9] QinFlow’s estimate based on its market research; the stated prices have not been confirmed by the manufacturer

DISCLAIMERS: The table above compares selected product attributes of buddy lite™ (a registered trademark of Belmont® Medical Technologies) with those of the Warrior configurations (manufactured by Quality In Flow, or QinFlow). Quality in Flow is not affiliated with Belmont® Medical Technologies. The benchmarking is based on publicly available information and the respective device’s Instructions for Use (IFU) / User Manual, where applicable. The buddy lite™ IFU / User Manual used for this study may not be the most updated version. Further, the IFUs / User Manuals used for this study may be changed by the manufacturer as the product evolves; QinFlow may not know about such updates and in any event does not undertake to update the information in the table upon such changes. Unless stated differently, the benchmark information was commissioned by QinFlow alone and not all data points were independently verified. Belmont® Medical Technologies was not part of this benchmark study. Use of the Warrior is subject to QinFlow’s terms of use. This benchmark study does not intend to cover all the differences between the devices. QinFlow does not guarantee that buddy lite™ will achieve the results or perform as mentioned in the benchmark study. QinFlow only warrants the information published in its IFUs (performance data points above are in accordance with the CE marks of each Warrior configuration; for FDA-cleared IFU, where applicable, contact QinFlow). Please contact Belmont® Medical Technologies for clarifications. 

Found an error? Please let us know at info@qinflow.com