PressSure™, including its air tube, is reusable. Clean it only when you determine it is necessary and in accordance with your organization’s infection-control policy.
1. Turn PressSure™ off and disconnect it from the pressure infusion bag.
2. Clean PressSure™ using 70% alcohol detergent or wipes.
3. Clean PressSure™ without allowing liquid penetrate into the charging port.
4. Allow PressSure™ to dry completely before charging, storing, or reusing.
5. Inspect the device and air tube after cleaning. Replace the tube whenever you deem necessary due to damage, contamination, or other operational reasons. There is no predefined routine replacement interval.
Note:
• Do not submerge PressSure™ or immerse the air tube connector.
• Do not sterilize or autoclave PressSure™. Do not use abrasive materials, sharp tools, or unapproved solvents.
• Do not spray cleaning solution directly into the display, buttons, charging port, or air tube connector.